Model Number 37800
Device Problems Migration or Expulsion of Device (1395); Impedance Problem (2950)
Patient Problems Pain (1994); Discomfort (2330)
Event Date 08/23/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) via a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator for gastric stimulation.It was reported that the patient had pain at the site of implant and the implant was moving around in the pocket, bothering the patient.It was also noted that impedance values were out of range, so the hcp turned the stimulator off.The patient was scheduled for a re-positioning of the battery for (b)(6) 2019.The issue was not resolved at the time of the report.No further complications were reported or anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a healthcare professional (hcp) via a manufacturer representative.It was reported that the physician repositioned the battery and the impedances normalized.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777 1200
MDR Report Key9007929
MDR Text Key160442246
Report Number3007566237-2019-01949
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2020
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/29/2019
Initial Date FDA Received09/12/2019
Supplement Dates Manufacturer Received09/23/2019
Supplement Dates FDA Received10/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age38 YR
-
-