Model Number 37800 |
Device Problems
Migration or Expulsion of Device (1395); Impedance Problem (2950)
|
Patient Problems
Pain (1994); Discomfort (2330)
|
Event Date 08/23/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare professional (hcp) via a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator for gastric stimulation.It was reported that the patient had pain at the site of implant and the implant was moving around in the pocket, bothering the patient.It was also noted that impedance values were out of range, so the hcp turned the stimulator off.The patient was scheduled for a re-positioning of the battery for (b)(6) 2019.The issue was not resolved at the time of the report.No further complications were reported or anticipated.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a healthcare professional (hcp) via a manufacturer representative.It was reported that the physician repositioned the battery and the impedances normalized.No further complications were reported/anticipated.
|
|