Model Number A3DR01 |
Device Problem
Reset Problem (3019)
|
Patient Problems
Muscle Stimulation (1412); Complaint, Ill-Defined (2331)
|
Event Date 08/22/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient has felt poorly since they were at their chiropractors two weeks ago and electrical therapy was used on their left shoulder.Upon interrogation, it was found that implantable pulse generator (ipg) had a full electrical reset.The patient symptoms are likely related to the vvi 65 pacing.The ipg remains in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that the patient experienced pectoral muscle stimulation.The implantable pulse generator (ipg) was programmed back to its original setting and symptoms resolved.
|
|