• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MALTA BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MALTA BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number MMC9609L
Device Problem Complete Blockage (1094)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/15/2019
Event Type  malfunction  
Manufacturer Narrative
Lot number - the customer reported a potentially associated lot number, 18k13t426.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a blood set had clots during priming.The reporter stated that when priming the set for a blood transfusion, large clots were noticed in the chamber of the set.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information :lot number provided: 18k13t426 : the device was received contaminated with blood.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional flow test was performed and liquid flows through appropriately.The reported problem was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - MALTA
marsa
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
marsa
MT  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9031413
MDR Text Key159334459
Report Number1416980-2019-05007
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMMC9609L
Device Lot Number18K13T426
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/16/2019
Initial Date FDA Received09/13/2019
Supplement Dates Manufacturer Received10/08/2019
Supplement Dates FDA Received10/16/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-