Model Number 2420-0500 |
Device Problem
Material Rupture (1546)
|
Patient Problem
No Information (3190)
|
Event Date 08/21/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.
|
|
Event Description
|
It was reported that the iv infusion pump began to alarm.When the nurse opened the device's door, the tubing's pump segment burst.
|
|
Event Description
|
It was reported that the iv infusion pump began to alarm.When the nurse opened the device's door and attempted to "massage" the tubing the pump segment split.
|
|
Manufacturer Narrative
|
Patient information requested, however not provided by the customer.The customer's report that the tubing's pump segment split was not confirmed, however a separation at the pump segment to the upper fitment was found.Visual inspection as received observed a separation between the silicone pump segment and upper fitment.Functional testing could not be performed due to the set's separation and obvious leaking that would occur due to the separation.Dimensional analysis was performed and the values measured were all within specification.The root cause of the separation could not be determined.
|
|
Search Alerts/Recalls
|