RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Loss of Power (1475); Vibration (1674)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2018.Date of report: 13sep2019.The manufacturer¿s international service technician replaced the power management (pm) pcba to address the reported problem.The unit was checked overall, run in tested, cleaned and functionally tested and no abnormality was confirmed.This mdr has been reassessed as reportable after a request from the fda on march 1, 2019 to review complaints from 29 nov 2017 to 29 nov 2018.As this has been reassessed, it will appear to be a late mdr.
|
|
Event Description
|
During periodic maintenance, the manufacturer¿s international service technician reported that the power of the light-emitting diode would turn off from vibration, such as a button operation.The manufacturer¿s international service technician reported there was no patient involvement when the event occurred.
|
|