This customer had a cystogram completed.There were several digital images taken during the fluoroscopy portion, and then an overhead x-ray after the procedure was completed.The technicians attempted to send the digital images to pacs.They were unable to get the digital images to transfer.After attempts made by a supervisor to resolve the issue to get the images to go (has been successful at times before), were unsuccessful, the supervisor contacted philips to assist.The representative made several attempts to resend the study, but the rep determined that the digital images were ultimately lost.There was the one overhead image that was sent to pacs and the radiologist was still able to provide a diagnosis.Biomedical engineering records: equipment passed preventative maintenance performed by philips on 3/18/19.Equipment passed medical physicist's fluoroscopy performance test summary on 5/16/19.
|