A customer in (b)(6) reported false negative results in association with the vidas® brahms procalcitonin assay ( lot: 1007056340 - expiry date: 03-jan-2020).On (b)(6) 2019, the customer performed a test with a sample from a (b)(6) newborn, which had a positive blood culture and high crp.The result obtained was < 0.05 ng/ml (negative) on the-a1 position of the minividas instrument.Following the clinical context, the customer tested the same tube on (b)(6) 2019 and obtained 40.03 ng/ml.The customer stated qcv results were conforming and there was no tube inversion during the test.The first result obtained has been reported to the physician, and no information regarding the patient's outcome was reported.There is no indication or report from the hospital or treating physician to biomérieux that the discrepant result led to any adverse event related to the patient's state of health.A biomérieux internal investigation will be initiated.
|
An internal biomérieux investigation was initiated for a customer report of obtaining a false positive result for a patient known to be negative when testing with the vidas® rubella igg ii assay (ref: (b)(4), lot: 1007220320) at position a1.The customer submitted the patient sample for the investigation.The investigation included a review of batch history records, testing of internal samples and the customer's returned patient sample with the customer's lot, and a review of vidas 3 instrument data.The analysis of the batch history records determined there was no anomaly during the manufacturing , control and packaging processes.A study of internal samples control charts was performed.Four internal samples from activity panel were analyzed.This analysis was carried out on all lots of vidas rub igg ii between lot: 191210-0 and lot: 200724-0 including customer lot: 200313-0.Results were within specifications.The customer's lot is in the trend of the other lots.Internal testing was performed on internal samples (negative and positive sera) and on the returned patient sample (b)(4).The return patient sample as well as four internal sera were tested on vidas 3 in automatic mode on retained kit, vidas rub igg ii batch: 1007220320 / 200313-0.The results obtained for the return sample are negative (2 ui/ml).The customer's anomaly was not reproduced internally.All results obtained for the four internal samples are in accordance with their expected specifications.The vidas 3 log files analysis showed that there was no issue on the instrument for the vidas rub igg ii assay regarding: the temperatures of the sections, the section pumps, the scanner head.Conclusion: the false positive result obtained by the client is an isolated anomaly, not reproduced internally and not explained according to our investigation results.According to the investigation, vidas rub igg ii lot: 1007220320 is performing as intended.
|