Model Number G136
Device Problems Fluid/Blood Leak (1250); Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a g136, the unit was leaking at all times and the handpiece is overheating; no injury resulted.
 
Manufacturer Narrative
Unit arrived with no filter in water supply line.Unknown how long it was run without a filter.Debris from unfiltered water damaged the water regulator causing it to leak.It appeared that unit was run with water control on handpiece cable turned all the way down to stop the leaking when not in use.Operating with the water control turned all the way down will cause the sterimate to overheat.Replaced water regulator and sterimate.Flushed water supply line and installed water filter.Calibrated and checked all functions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON PLUS ULTRASONIC SCALER WITH TAP-ON TECHNOLOGY
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
MDR Report Key9073478
MDR Text Key186607461
Report Number2424472-2019-00138
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 10/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG136
Device Catalogue Number8184001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/16/2019
Supplement Dates Manufacturer Received09/12/2019
Supplement Dates FDA Received10/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-