• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM Back to Search Results
Model Number 6379
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 08/23/2019
Event Type  malfunction  
Manufacturer Narrative
The fsr replaced a blown fuse.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer for further evaluation.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the charging green light emitting diode (led) was not illuminating.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed the filament inside the fuse to be charred and does not have continuity when tested.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9075420
MDR Text Key159611409
Report Number1828100-2019-00499
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number6379
Device Catalogue Number6379
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/23/2019
Initial Date FDA Received09/17/2019
Supplement Dates Manufacturer Received12/20/2019
Supplement Dates FDA Received12/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-