There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa.(510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative in charge found out that the patient pump 2 was stuck.The part was replaced and the issue solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Pictures of the affected parts has been provided as evidence.The analysis of the pictures revealed that the bearing damage is visible and this had led to the reported error code outbreak.A review of the dhr could not identify any deviations or non conformities relevant to the issue.
|