Model Number 16-02-82
Device Problem Power Problem (3010)
Patient Problem No Patient Involvement (2645)
Event Date 08/22/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the issue.The technician stated that there was a power surge and it shorted out the power supply.The power supply board was replaced and a new two pins plug was installed.The issue was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t was not turning on during setup.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: a visual inspection of the affected part was performed and revealed burn marks on a connector and terminals.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key9087232
MDR Text Key178158772
Report Number9611109-2019-00737
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-82
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/22/2019
Initial Date FDA Received09/19/2019
Supplement Dates Manufacturer Received10/08/2019
Supplement Dates FDA Received11/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-