The customer¿s report of breakage was confirmed.However, the breakage occurred on the smartsite component of the vial adapter and not on the syringe.The syringe and vad were visually inspected for obvious damage such as incomplete bonding engagements, cracks/fractures, crazing, breaks or other damage and anomalies.Visual inspection of the set noted the white female luer adapter (fla), of the vad¿s smartsite component that is welded to the clear smartsite body was broken off from the body of the smartsite.Further inspection of the fla end found a remnant ring of the clear smartsite body remaining within (still welded to) the fla.The corresponding surface areas of the breakage were white with uneven and jagged edges (not a ¿clean¿ break).The uneven breakage and white areas are an indication of stress from the application of external force.Examination of the fla under magnification revealed that the threads of the smartsite were turned upwards indicating possible over-tightening to the texium mating component.No other anomalies or evidence of damage were observed.Functional testing was not performed due to the obvious damage.The breakage was caused by an external force.The root cause of the external force was not identified.
|