The patient was undergoing a thrombectomy procedure in the tibiofibular trunk using an indigo system cat 6 aspiration catheter (cat6).During the procedure, while attempting to insert a cat6 into a non-penumbra sheath using the peel away sheath, the tip of the cat6 became came apart and unraveled; therefore, it was removed.The procedure was completed using another cat6 and the same sheath.There was no report of an adverse effect to the patient.
|
Results: the cat6 had multiple kinks, a stretch and ovalizations from approximately 118.0 ¿ 136.0 cm.The total length of the cat6 was approximately 136.0 cm.Conclusions: evaluation of the returned cat6 revealed multiple kinks, a stretch and ovalizations along the distal shaft.If the peel away sheath provided with the cat6 is not properly utilized during insertion of the cat6 into a parent device, damages such as these may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|