• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 6; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 6; DXE Back to Search Results
Catalog Number CAT6KIT
Device Problem Unraveled Material (1664)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/21/2019
Event Type  malfunction  
Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event Description
The patient was undergoing a thrombectomy procedure in the tibiofibular trunk using an indigo system cat 6 aspiration catheter (cat6).During the procedure, while attempting to insert a cat6 into a non-penumbra sheath using the peel away sheath, the tip of the cat6 became came apart and unraveled; therefore, it was removed.The procedure was completed using another cat6 and the same sheath.There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
Results: the cat6 had multiple kinks, a stretch and ovalizations from approximately 118.0 ¿ 136.0 cm.The total length of the cat6 was approximately 136.0 cm.Conclusions: evaluation of the returned cat6 revealed multiple kinks, a stretch and ovalizations along the distal shaft.If the peel away sheath provided with the cat6 is not properly utilized during insertion of the cat6 into a parent device, damages such as these may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 6
Type of Device
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9090148
MDR Text Key159315734
Report Number3005168196-2019-01770
Device Sequence Number1
Product Code DXE
UDI-Device Identifier00815948020405
UDI-Public00815948020405
Combination Product (y/n)Y
PMA/PMN Number
K160533
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/01/2005,08/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/19/2022
Device Catalogue NumberCAT6KIT
Device Lot NumberF91123
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2019
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 08/21/2019
Initial Date FDA Received09/19/2019
Supplement Dates Manufacturer Received09/23/2019
Supplement Dates FDA Received10/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age79 YR
-
-