MAUDE Adverse Event Report: MERIT MEDICAL SINGAPORE PTE LTD. SAFEDRAW BLOOD SAMPLING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SINGAPORE PTE LTD. SAFEDRAW BLOOD SAMPLING SYSTEM Back to Search Results
Catalog Number 689069/B
Device Problem Material Separation (1562)
Patient Problem No Patient Involvement (2645)
Event Date 08/26/2019
Event Type  malfunction  
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that the chamber is detaching from the spike with minimal force during testing.The device was not used on a patient.
 
Manufacturer Narrative
The suspect device has been returned for evaluation.The product was examined visually.The complaint is confirmed.The root cause is attributed to the manufacturing process.A search of the complaint database was performed and two similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.Corrective actions are in progress.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDRAW BLOOD SAMPLING SYSTEM
Type of Device
BLOOD SAMPLING SYSTEM
Manufacturer (Section D)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
76892 6
SN  768926
MDR Report Key9098784
MDR Text Key159576959
Report Number8020616-2019-00103
Device Sequence Number1
Product Code DSK
Combination Product (y/n)N
PMA/PMN Number
K885235
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 08/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number689069/B
Device Lot NumberC1446687
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2019
Initial Date Manufacturer Received 08/26/2019
Initial Date FDA Received09/20/2019
Supplement Dates Manufacturer Received10/14/2019
Supplement Dates FDA Received10/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-