• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Device Problems Crack (1135); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 09/23/2019.
 
Event Description
It was reported that the display became detached from the unit and that the cover was cracked.The device was in use at the time of the event; however, there was no patient harm.
 
Manufacturer Narrative
Date received by mfr: 25nov2019.The manufacturer's field service engineer (fse) performed troubleshooting and confirmed the reported issue.The fse replaced the top cover assembly, display cover assembly, keypad membrane, power cord strain relief clip and rubber feet and the unit was repaired.No parts were returned to failure investigation, therefore, the root cause of the reported issue could not be determined.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9104117
MDR Text Key187588585
Report Number2031642-2019-09776
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/13/2019
Initial Date FDA Received09/23/2019
Supplement Dates Manufacturer Received09/13/2019
Supplement Dates FDA Received11/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK AND HUMIDIFIER.
-
-