ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problems Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that a couple of doctors have noted that when they use cartridges during intraocular lens (iol) implant procedures there seems to be a strip that comes off and ends up in the eye.Both doctors have noted that the strip is quite difficult to remove from the eye and that it seems to be firm and plastic like.Additional information was provided that the event happens when the cartridge is used with a competitors iol and or a competitors toric iol.We have used these cartridges with these lenses in the past with no issue.The procedures were completed.The plastic piece that was in the eye after lens insertion was removed by irrigation / aspiration and or removed with a forcep.The event has occurred with multiple patients.No harm and the iols remained intact.There are three medical device reports associated with the cartridges reported.This report is for one of three lot numbers reported with an unknown date of event.
 
Manufacturer Narrative
Evaluation summary: the customer indicated the use of non-qualified associated products.The root cause is most likely a failure to follow the directions for use (dfu).The account used a non-qualified lens/cartridge combination.The use of non-qualified combinations may result in delivery issues and/or damage.Customer has indicated the use of different manufacturer's iols.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key9105671
MDR Text Key165635881
Report Number1119421-2019-01551
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number8065977763
Device Lot Number32683471
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/28/2019
Initial Date FDA Received09/23/2019
Supplement Dates Manufacturer Received10/21/2019
Supplement Dates FDA Received11/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ABBOTT MEDICAL OPTICS IOL AND OR TORIC IOL; HEALON GV; MONARCH II LOADING; UNSPECIFIED ALCON HANDPIECE; UNSPECIFIED DISCOVISC; UNSPECIFIED PROVISC
-
-