It was reported that the 3 months after stent placement, it was found that the stent was found to be broken.Through the attached x-ray photo, it is confirmed that the stent was fractured.It was observed that the foreign substance was attached to removed stent wire.It was confirmed from the device history record that device had been manufactured with no significant issue and passed all the inspections successfully.Fracture can be occurred by other company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.Duodenal structure where stent was implanted is curvy.Stent can be frequently pressured due to patient's lesion status and fracture be possible.However, it is hard to identify the exact root cause since it is hard to reconstruct the situation at the time of procedure.Based on that it is confirmed that the foreign substance was attached on the stent wire in photo, it is assumed that the wire was somewhat weakened due to the patient lesion's peristalses and pressure and foreign substance such as foods, body fluids and so on, resulted in stent fracture.Taewoong medical has managed the risk of stent fracture through the fmea, and it is confirmed that it is occurring within acceptable criteria because the stent fracture complaint is 1 case until the present in 2019.This suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
|