• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT; SCALER, ULTRASONIC Back to Search Results
Catalog Number 80395
Device Problem Overheating of Device (1437)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Further investigation showed some material got inside the insert hole clogging the insert, and as a result, the insert was getting hot because of water flow restriction.This band was installed by the customer.The issue was due to material debris from the band installed by the user.No manufacturing problem was identified.Multiple unsuccessful attempts were made to obtain the patient outcome.
 
Event Description
While the customer was using a 30k fsi-sli-10s insert, it was running hot.The event outcome is unknown as of this mdr evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
karl nittinger
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key9127018
MDR Text Key186714832
Report Number2424472-2019-00148
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 09/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80395
Device Lot Number00031368
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/27/2019
Initial Date FDA Received09/26/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-