There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Further investigation showed some material got inside the insert hole clogging the insert, and as a result, the insert was getting hot because of water flow restriction.This band was installed by the customer.The issue was due to material debris from the band installed by the user.No manufacturing problem was identified.Multiple unsuccessful attempts were made to obtain the patient outcome.
|