Model Number 10-80-00 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/12/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).Livanova (b)(4) initiated an investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 roller pump displayed an error message at the installation phase.Two board were replaced but the issue still persisted.The pump head was replaced and the pump was working again.There was no patient involvement.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: deviation could be reproduced during the investigation at the manufacturer site.A defective electronic component (resolver) of the pump head was identified as the root cause of the reported event.
|
|