The customer reported that in the last weeks, too low ica values were measured on the abl800 flex analyzer.The described error resulted in several patients in the operating room being treated with intravenous calcium, but no permanent damage was reported.Measurements on the affected device (serial no.(b)(4)) compared with two other devices are shown in the table below.Abl800 sn: (b)(4).Abl800 sn: (b)(4).Abl800 sn: (b)(4).(b)(6).
|
The following additional information has been received from the customer: " since (b)(6), 153 measurements have been performed with the device, where ca++ has been measured in 40 analyses in the reference range, 5 analyses above the reference range and 108 analyses below the reference range.Compared to the previous months, these are a striking number of measurements where the ca++ is below the reference range.Only in one patient did we have a measurement on another device in the same period, which showed a ca++ in the reference range at initial ca++ below the reference range during the control measurement.Both measuring chambers were inspected by the field service engineer and damages was observed.The cylinders were not fixed well anymore.The ca++ and reference electrodes including used membranes will be returned to radiometer for investigation.
|