It was reported that during a reverse surgery arthroplasty, while reaming the glenoid, the resident in the case broke pilot tip on the starter reamer.The broken tip was removed from the patient and this was verified via fluoroscopy.The patient was stable as they left the or.
|
Section h10: (h3) based on capa2017-12, the broken device reported was likely the result of applying a bending moment to the reamer during use, which led to brittle fracture of the pilot tip feature.(h6) evaluation codes: 2199, 1069.(h7) recall.(h9) if action reported to fda under 21 usc 360i(f), list correction/removal reporting number: z-2663-2017, z-2664-2017, z-2665-2017, z-2666-2017, z-2667-2017, z-2668-2017.
|