BUNNELL, INCORPORATED BUNNELL; VENTILATOR, HIGH FREQUENCY Back to Search Results
Device Problems Device Alarm System (1012); Loss of Power (1475)
Patient Problem Low Oxygen Saturation (2477)
Event Date 07/16/2019
Event Type  malfunction  
Event Description
Bunnell jet alarm going off respiratory therapist (rt) went over to check found jet stopped and reset to default settings, infant desaturating to low 80's.There was no noted glitch or flicker in electricity.Checked circuit to make sure all connections were secure, tight and nothing was kinked.Reset appropriate settings and restarted jet.This happened 2x's more and error code 05 was seen on jet screen, and then rt proceeded to switch out jet.Error code 05 is an electrical irregularity has disrupted the microprocessor, corrective action was to power off life pule for 3-4 mins and then power back on.Rt opted to change out device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BUNNELL
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
BUNNELL, INCORPORATED
436 lawndale dr.
s salt lake UT 84115
MDR Report Key9145073
MDR Text Key160890060
Report Number9145073
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 09/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Other
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA09/13/2019
Event Location Hospital
Date Report to Manufacturer10/02/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age18 DA
-
-