Model Number 3100 A |
Device Problem
Defective Alarm (1014)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/03/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Any additional information provided by the customer will be included in a follow up report.The healthcare facility attempted to replace the alarm board on this device.However, the reported issue was not resolved.At this time, vyaire has not received the suspect device/component for evaluation.
|
|
Event Description
|
It was reported to vyaire medical that the 3100 a unit power failure alarm will not alarm.The customer confirmed that there was no patient involvement associated with the reported issue.
|
|
Manufacturer Narrative
|
The vyaire field service representative (fsr) was able to replicate the reported problem.The field service technician was able to resolve the issue by re-seating the 9 volt battery and all alarm cables and connectors.Performed pt calibration and performance check, all passed per 3100a ventilator label.
|
|
Search Alerts/Recalls
|