• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE ¿ ROUND LAKE EM 2400, DISPLAY MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE ¿ ROUND LAKE EM 2400, DISPLAY MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number EXACTA24DY
Device Problems Improper or Incorrect Procedure or Method (2017); Inaccurate Information (4051)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/04/2019
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the display of an exactamix automated compounding device had misused lipid products.Customer stated "the compounder had been set up incorrectly and their technicians applied the smof lipid barcode labels to their intralipid containers.This caused the compounder to deliver intralipid in place of smof lipid during formula runs".The issue had occurred during compounding; however, it was reported there was patient involvement.Therefore, the bag was infused to a patient.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: it was reported the customer misused the lipid products by setting up the device incorrectly.The technicians applied the smof lipid barcode labels to their intralipid containers.Use errors and proper user instructions are addressed in ¿extamix operator manual¿, which is included with every extamix device.The guide instructs the user to make sure the user labels the inlet with the appropriate, numbered bar code label that was packaged with the valve set.It also tells the user to check that the correct inlet is attached to the correct ingredient and port, as well as, labeled with the correct bar code label.A review of the label for the product family will be conducted.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM 2400, DISPLAY MODULE
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE ¿ ROUND LAKE
round lake IL
Manufacturer (Section G)
BAXTER HEALTHCARE - ROUND LAKE
2575 university ave w
round lake IL 60073
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9147687
MDR Text Key161222474
Report Number1416980-2019-05391
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberEXACTA24DY
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/04/2019
Initial Date FDA Received10/02/2019
Supplement Dates Manufacturer Received09/04/2019
Supplement Dates FDA Received10/04/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-