The associated complaint device was returned for evaluation.A visual inspection of the returned stem inserter shows the tip is broken off.The complaint description states the tip is stuck in the stem.The stem was not returned for evaluation.The device was manufactured in 2002 and it shows signs of extreme wear usage.A clinical evaluation was conducted and based on the limited information provided, the root cause of the inserter tip breakage cannot be determined.The material, 17-4ph is not an implantable alloy; therefore, long-term implantation data is not available.Since the inserter tip was left wedged inside the femoral component, migration is unlikely.The impact to the patient is the retained non-implantable fragment of the externally communicating device, possible local irritation and/or discomfort, and additional radiological imaging/exposure cannot be completely ruled out.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal any manufacturing abnormalities that could have caused or contributed to the reported incident.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
|