Interrogation of the device revealed the device was at elective replacement indicator (eri) when received.A longevity calculation was performed based upon programmed settings and usage data.The calculations showed the battery depletion was normal.A device evaluation was performed, and no sources of high current were noted.The telemetry, pacing, sensing, impedance, hv output, hv shock, and patient notifier were tested on the bench.No anomaly was detected.The pacing output issue observed in the field was believed to be due to electrocautery exposure during explant procedure.
|