Catalog Number 8065990739
Device Problem Patient-Device Incompatibility (2682)
Patient Problems Irritation (1941); Red Eye(s) (2038)
Event Date 09/23/2019
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The device history records (dhr) for the device was reviewed.The associated device was released based on company acceptance criteria.No udi required due to this device being out of production prior to the (b)(4) 2014 udi regulation date.The manufacturer internal reference number is: (b)(4).
 
Event Description
An optometrist reported a patient that complained of redness and irritation in the right eye four days post lasik.The patient noticed a white spot near the pupil.New information received stated the patient was seen for two additional follow up visits and the infiltrate is no longer presumed infectious.
 
Manufacturer Narrative
A review of the technical service onsite history showed no abnormalities that could have contributed to this event: laser was successfully verified prior to and after the day of treatment.Logfile review shows no abnormalities, all laser system functions were within specifications at this day.No technical root cause was identified as the product was found to be within specifications.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
MDR Report Key9157715
MDR Text Key162628524
Report Number3003288808-2019-00986
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 12/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065990739
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/23/2019
Initial Date FDA Received10/04/2019
Supplement Dates Manufacturer Received11/14/2019
Supplement Dates FDA Received12/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age26 YR
-
-