Catalog Number 8065990739 |
Device Problem
Patient-Device Incompatibility (2682)
|
Patient Problems
Irritation (1941); Red Eye(s) (2038)
|
Event Date 09/23/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The device history records (dhr) for the device was reviewed.The associated device was released based on company acceptance criteria.No udi required due to this device being out of production prior to the (b)(4) 2014 udi regulation date.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
An optometrist reported a patient that complained of redness and irritation in the right eye four days post lasik.The patient noticed a white spot near the pupil.New information received stated the patient was seen for two additional follow up visits and the infiltrate is no longer presumed infectious.
|
|
Manufacturer Narrative
|
A review of the technical service onsite history showed no abnormalities that could have contributed to this event: laser was successfully verified prior to and after the day of treatment.Logfile review shows no abnormalities, all laser system functions were within specifications at this day.No technical root cause was identified as the product was found to be within specifications.The manufacturer internal reference number is: (b)(4).
|
|