Model Number MSERIES CCT |
Device Problems
Unable to Obtain Readings (1516); Unexpected Shutdown (4019)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Additional pro code: mkj.Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during functional testing, the device displayed a "poor pad contact" message and inappropriately shut down.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical (b)(4); the reported malfunction for device shut down was duplicated and the ac power cord was replaced to remedy the problem.The device was recertified and returned to the customer.The reported malfunction for device displayed "poor pad contact" was evaluated and communication with the customer indicated there were no electrode pads attached at the time.This is normal operation of the device if there are no pads connected to the device.No trend is associated with reports of this type.
|
|