• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS ADVANCE XP SLING SYSTEM; MESH SURGICAL FOR STRESS URINARY INCONTINENCE MALE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS ADVANCE XP SLING SYSTEM; MESH SURGICAL FOR STRESS URINARY INCONTINENCE MALE Back to Search Results
Model Number 720163-01
Device Problems Premature Activation (1484); Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/24/2019
Event Type  Injury  
Event Description
It was reported that during a sling procedure the mesh came unattached with the trocar before getting to the skin surface when the physician tried to pull the sling arms up.One side of the mesh would not reattach to the trocar.The physician used a silk and followed the hole on the sides of the cone connector and then went through the bottom out the top of the cone, used a surgeons knot to adhere to the trocar.A hammer/ richardson tool with a free keith needle was used to go through the cone from bottom to top.The suture worked great.It was also stated that all was good.It was further reported that the patient did not experience any additional adverse events and the sling connection issue did not extend the procedure time significantly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS ADVANCE XP SLING SYSTEM
Type of Device
MESH SURGICAL FOR STRESS URINARY INCONTINENCE MALE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
100 boston scientific way
marlborough MA 01752 1242
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
alyson harris
10700 bren road w
minnetonka, MN 55343
4083953452
MDR Report Key9191421
MDR Text Key163159904
Report Number2183959-2019-66840
Device Sequence Number1
Product Code OTM
UDI-Device Identifier00878953005560
UDI-Public00878953005560
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/12/2019
Device Model Number720163-01
Device Catalogue Number720163-01
Device Lot Number0023196988
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/24/2019
Initial Date FDA Received10/15/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/16/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age72 YR
-
-