Overall investigation summary: customer called guerbet service reporting that their hydravision urology system was unable to produce fluoro during a case.The customer stated that they tried cycling the power on the computer, generator and the table, but were still not able to produce fluoro.Guerbet service suggested transferring the call to guerbet's technical support so that immediate assistance could be provided.However, the customer declined, asking that a field service engineer(fse) be dispatched to check the system.The customer then moved the patient to another table and completed the case using portable fluoroscopy.Guerbet's the fse went on site to confirm the complaint.In addition to power-cycling the computer, generator and table, he also power-cycled the thales imaging processor and detector, after which the system booted up normally and was able to produce fluoro.The fse then verified proper operation using service checklist qssrwi4.3 and returned the system to full service.A review of cts shows no similar issues reported on this unit.Root/probable cause code: equipment/instrument - software.Root / probable cause summary: the thales imaging processor required a reboot.No further investigation needed at this time.Qa will continue to monitor and trend for similar issues.No capa at this time, these trends and issues are reported on during quality metrics review and during the management review meetings to consider input for corrective action.Disposition summary: unit remained in full service.
|
This incident was reported on (b)(6) 2019, as customer called guerbet service reporting that their hydravision urology system was unable to produce fluoro during a case.The customer stated that they tried cycling the power on the computer, generator and the table, but were still not able to produce fluoro.
|