• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVE DIVISION AORTIC VALVE GRAFT; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVE DIVISION AORTIC VALVE GRAFT; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 502AG23
Device Problems Perivalvular Leak (1457); Insufficient Information (3190)
Patient Problems Aortic Regurgitation (1716); Blood Loss (2597)
Event Date 09/19/2019
Event Type  Injury  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that immediately post implant of this aortic valved graft, the graft was explanted and replaced with a non-medtronic device.The reason for explant was due to blood seepage from the graft and as a result the physician chose to explant and replace it.No additional adverse patient effects were reported. .
 
Manufacturer Narrative
Medtronic received additional information that there was also moderate regurgitation present.No additional adverse patient effects were reported.  medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Conclusion: based on the reported information and device investigation, the reported regurgitation was most likely paravalvular leak which can be caused by a variety of factors, including valve positioning due to implant procedure, patient anatomy, or presence of pre-existing patient conditions, and a conclusive cause could not be determined from the limited information available.The device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.H6: coding updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC VALVE GRAFT
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
3800 annapolis lane
minneapolis MN 55447
MDR Report Key9196920
MDR Text Key162631979
Report Number3008592544-2019-00050
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
PMA/PMN Number
P990046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/22/2022
Device Model Number502AG23
Device Catalogue Number502AG23
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/19/2019
Initial Date FDA Received10/16/2019
Supplement Dates Manufacturer Received10/18/2019
02/26/2020
Supplement Dates FDA Received10/21/2019
02/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age31 YR
-
-