MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT ALUMINA INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT ALUMINA INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE Back to Search Results
Catalog Number 625-0T-32F
Device Problems Crack (1135); Fracture (1260)
Patient Problem Injury (2348)
Event Date 09/19/2019
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event Description
It was reported that an alleged broken ceramic head and liner was retrieved from a patient.
 
Manufacturer Narrative
An event regarding damage involving a trident liner was reported.The event was confirmed through material analysis of the returned device.Method & results: product evaluation and results: visual inspection was performed as part of the material analysis report (mar), dated 20 november 2019.This inspection indicated edge cracking was observed on the alumina insert due to contact against the stem.Metal transfer markings were observed on the alumina insert consistent with contact against the stem.Edge cracking was observed on the alumina insert consistent with contact against the stem.A material analysis has been performed.The report concluded: the alumina head was returned fractured in pieces and no evidence of a continuous metal transfer ring was observed.Impingement from the stem and damage consistent with contact against a hard object was observed on the distal rim of the titanium sleeve.Edge cracking was observed on the alumina insert due to contact against the stem in addition to metal transfer markings.Damage consistent with the implantation/explantation process was observed on the anterior and posterior surfaces of the stem.Damage consistent with impingement on the shell was observed on the inferior surface of the stem.Debris was also observed on the anterior, posterior, and inferior surfaces of the stem and elemental analysis confirmed it to be consistent with an oxide, base material, biological material and the decontamination process.Debris was observed on the proximal surface of the shell and elemental analysis confirmed it to be consistent with an oxide, biological material, the base material, and an unknown material.Elemental analysis confirmed that the shell and stem are consistent with an astm f136 alloy and an astm f1586 alloy, respectively.Based on the given information, no identifiable material or manufacturing discrepancies were observed on the surfaces examined.Clinician review: no medical records were received for review with a clinical consultant.Product history review: indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: the material analysis report concluded.The alumina head was returned fractured in pieces and no evidence of a continuous metal transfer ring was observed.Impingement from the stem and damage consistent with contact against a hard object was observed on the distal rim of the titanium sleeve.Edge cracking was observed on the alumina insert due to contact against the stem in addition to metal transfer markings.Damage consistent with the implantation/explantation process was observed on the anterior and posterior surfaces of the stem.Damage consistent with impingement on the shell was observed on the inferior surface of the stem.Debris was also observed on the anterior, posterior, and inferior surfaces of the stem and elemental analysis confirmed it to be consistent with an oxide, base material, biological material and the decontamination process.Debris was observed on the proximal surface of the shell and elemental analysis confirmed it to be consistent with an oxide, biological material, the base material, and an unknown material.Elemental analysis confirmed that the shell and stem are consistent with an astm f136 alloy and an astm f1586 alloy, respectively.Based on the given information, no identifiable material or manufacturing discrepancies were observed on the surfaces examined.The exact cause of the event cannot be confirmed as insufficient information was provided.Further information such as pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
It was reported that an alleged broken ceramic head and liner was retrieved from a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT ALUMINA INSERT
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9197359
MDR Text Key170056749
Report Number0002249697-2019-03514
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
PMA/PMN Number
P000013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Catalogue Number625-0T-32F
Device Lot Number45151705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2019
Initial Date Manufacturer Received 09/19/2019
Initial Date FDA Received10/16/2019
Supplement Dates Manufacturer Received12/02/2019
Supplement Dates FDA Received12/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-