Model Number 1050044 |
Device Problems
Fluid/Blood Leak (1250); Mechanical Problem (1384)
|
Patient Problem
Patient Problem/Medical Problem (2688)
|
Event Date 09/27/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Pma/510k: reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.K111959.When additional information becomes available, a follow up report will be submitted.
|
|
Event Description
|
It was reported the ampule leaked.The reporter indicated that when opening the package, the leakage from the ampule was already caused, therefore the product was not able to be used.
|
|
Manufacturer Narrative
|
Investigation: samples received: 1 open pouch, ampoule unopened with leakage signs.Analysis and results: there are previous complaints of this code batch of which we manufactured and distributed in the market (b)(4) units.There are no units in stock in b.Braun surgical warehouse.We have received an open pouch (closed ampoule) showing ampoule leakage.The ampoule received has been optically evaluated and a defect in the sealing bar of the ampoule was found.The leakage of the glue occurs at this point as can be seen in the enclosed picture.Reviewed the batch manufacturing record, this product had an incidence not related to this issue and was released fulfilling b.Braun surgical specifications.Final conclusion: taking into account that the results of the sample received do not fulfil b.Braun surgical specifications, and that there are several previous complaints for the same defect, we conclude that the complaint is confirmed by evidence of failure in the sample received.Actions on distributed product of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/proventative actions: a capa has been initiated.
|
|
Search Alerts/Recalls
|
|