BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number 8900
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2019
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Date of event is estimated based on the aware date of (b)(6) 2019.Returned product consisted of an ffr comet pressure wire not connected to the occ cable.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed 1 kink located at the 43.5cm from the tip.The coating was peeled at 43.5cm.The tip showed a bend.The collet was stuck inside of the cap.The wire was inserted into the occ handle and tested for signal issues.The signal was present as designed.The collet was removed from the cap and the cap was analyzed.The cap showed damage inside where the collect was stuck and would not open to release the wire.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Because there was no evidence of any product quality deficiencies, it was considered likely that the kink, tip damage and the coating issues were attributable to handling.
 
Event Description
Reportable based on analysis completed on 26sep2019.It was reported that the guidewire could not be disconnected.During a ffr procedure, the comet guidewire could not be disconnected from the handle as the wire was stuck.The guidewire was removed from the patient and the procedure was successfully completed with a different device without issue or patient injury.However, returned device analysis revealed coating peeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9205969
MDR Text Key164328181
Report Number2134265-2019-12447
Device Sequence Number1
Product Code DXO
UDI-Device Identifier08714729904397
UDI-Public08714729904397
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K151610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/04/2021
Device Model Number8900
Device Catalogue Number8900
Device Lot Number0023435725
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/26/2019
Initial Date FDA Received10/17/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-