Age at time of event: 18 years or older.Date of event is estimated based on the aware date of (b)(6) 2019.Returned product consisted of an ffr comet pressure wire not connected to the occ cable.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed 1 kink located at the 43.5cm from the tip.The coating was peeled at 43.5cm.The tip showed a bend.The collet was stuck inside of the cap.The wire was inserted into the occ handle and tested for signal issues.The signal was present as designed.The collet was removed from the cap and the cap was analyzed.The cap showed damage inside where the collect was stuck and would not open to release the wire.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Because there was no evidence of any product quality deficiencies, it was considered likely that the kink, tip damage and the coating issues were attributable to handling.
|
Reportable based on analysis completed on 26sep2019.It was reported that the guidewire could not be disconnected.During a ffr procedure, the comet guidewire could not be disconnected from the handle as the wire was stuck.The guidewire was removed from the patient and the procedure was successfully completed with a different device without issue or patient injury.However, returned device analysis revealed coating peeling.
|