• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500; BLOOD GAS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500; BLOOD GAS MONITOR Back to Search Results
Model Number CDI510H
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during out of box, there was a leak from the shunt sensor.No patient involvement.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on october 18, 2019.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 213, 67 ).Method code #1: 10 - testing of actual/suspected device.Method code #2: 11 - testing of device from same lot/batch retained by manufacturer.Method code #3: 3331 - analysis of production records.Results code: 213 - no device problem found.Conclusions code: 67 - no problem detected.The returned sample was visually inspected and there were no anomalies noted.It was then leak tested by connecting with the calibrated manometer, submerged into a water bath, and pressurized up to 1030 mmhg, and no leaks were noted.The large blue cap was then loosened and retightened with a calibrated torque wrench to production specification.The unit was then pressurized with air, submerged into a water bath, and pressurized up to 1030 mmhg, and no leaks were noted.A retention sample from the same lot number was visually inspected and confirmed to have no traces of buffer on the outside of the unit or inside the pouch.The retention sample was then pressurized with air up to 1030 mmhg, submerged in a water bath, and observed for any leaks.No leaks were noted on the retention sample.The complaint was not confirmed, and a definitive root cause was not able to be determined.It is possible that the buffer solution was not completely wiped off the product after buffer fill during manufacturing.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNT SENSOR SYS500
Type of Device
BLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key9210940
MDR Text Key164343212
Report Number1124841-2019-00286
Device Sequence Number1
Product Code DRY
UDI-Device Identifier00699753160767
UDI-Public(01)00699753160767
Combination Product (y/n)N
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Model NumberCDI510H
Device Catalogue NumberN/A
Device Lot NumberXE29A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/02/2019
Initial Date FDA Received10/18/2019
Supplement Dates Manufacturer Received11/26/2019
Supplement Dates FDA Received11/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-