It was reported that during the implant of this implantable cardioverter defibrillator (icd), shock impedance measurements of greater than 200 ohms were noted.A code 1005 was also observed, indicating an open circuit condition was detected during shock delivery.It was noted that the physician attempted to use a covering over the device due to a suspected metal allergy for the patient.A different lead was implanted with this same device with no further reported issues.No adverse patient effects were reported.
|