MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS U411 TEST SYSTEM; AUTOMATED URINALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS U411 TEST SYSTEM; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Model Number U411
Device Problem Patient Data Problem (3197)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The initial reporter stated that results were assigned to an incorrect sample when running samples on the cobas u 411 analyzer.Results for one sample were assigned to a different sample identifier.Error messages also appeared on the analyzer when this happened.The issue was resolved after the customer's laboratory information system was disconnected from the analyzer.
 
Manufacturer Narrative
No incorrect results were reported outside of the laboratory.The host was reconnected to the instrument and it is working without incident.No further information was provided.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS U411 TEST SYSTEM
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key9214169
MDR Text Key205986917
Report Number1823260-2019-03799
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
PMA/PMN Number
K093555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberU411
Device Catalogue Number04906969001
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/02/2019
Initial Date FDA Received10/21/2019
Supplement Dates Manufacturer Received10/02/2019
Supplement Dates FDA Received12/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-