• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA HP UNI TIB TRAY SZ3 LMRL; EARLY INTERVENTION : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA HP UNI TIB TRAY SZ3 LMRL; EARLY INTERVENTION : KNEE TIBIAL TRAY Back to Search Results
Model Number 1024-51-300
Device Problems Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993); Noise, Audible (3273)
Patient Problems Adhesion(s) (1695); Pain (1994); Discomfort (2330); No Code Available (3191)
Event Date 10/01/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the patient had clicking and some pain over time.The surgeon performed an arthroscopy in the past and noticed the uni poly was lifting up in the front and he was concerned that it could be causing discomfort to the patient.The surgeon would like a letter back to him responding to what was causing the poly to be lifting up.He was wondering if the poly was out of spec.To the tray that was originally implanted.Doi: (b)(6) 2016; dor: (b)(6) 2019; left knee.
 
Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : null.Device history batch : null.Device history review : null.
 
Event Description
Medical records reviewed.On (b)(6) 2016, patient received a medial left knee unicompartmental arthroplasty, to address medial compartment osteoarthritis.There was no reported complication.On (b)(6) 2018, patient's left knee was revised to address pain and suspected implant loosening.It was determined though, that there was no implant loosening--both femur and tibial tray components were well-fixed.Rather, it was identified that the tibial insert component locking mechanism had partially failed, with the insert moving in the tray.This insert component was revised, and scar tissue excised.
 
Manufacturer Narrative
Product complaint#: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  h10 additional narrative: corrected: d7 (this is being retracted as the device was not revised).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. .
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary :no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot :the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI TIB TRAY SZ3 LMRL
Type of Device
EARLY INTERVENTION : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 46581-0988
6107428552
MDR Report Key9221439
MDR Text Key169592121
Report Number1818910-2019-111277
Device Sequence Number1
Product Code HRY
UDI-Device Identifier10603295002338
UDI-Public10603295002338
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1024-51-300
Device Catalogue Number102451300
Device Lot Number2266
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/01/2019
Initial Date FDA Received10/22/2019
Supplement Dates Manufacturer Received11/20/2019
03/23/2020
09/07/2022
Supplement Dates FDA Received11/26/2019
04/03/2020
09/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
COMPETITOR BONE CEMENT.; SIGMA HP UNI FEMORAL SZ4 LM/RL.; SIGMA HP UNI INS SZ3 7MM LM/RL.
Patient Outcome(s) Required Intervention;
Patient Age59 YR
Patient SexMale
-
-