The initial reporter received questionable elecsys vitamin b12 immunoassay results for two patient sample from cobas e 801 analyzer serial number (b)(4).Patient 1 initial results were 98.9 pmol/l and 139 pmol/l.On (b)(6) 2019, the repeat result from a siemens analyzer was 275 pmol/l.Patient 2 initial results were 92.3 pmol/l and 95.6 pmol/l.On (b)(6) 2019, the repeat result from a siemens analyzer was 159 pmol/l.The results were reported outside of the laboratory and were questioned by the clinician.
|
The calibration data is acceptable, and the customer¿s qc data is within their ranges.Based on this, a general reagent issue can be excluded.Assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used and, differences in reference materials/methods and the standardization methodology used.The investigation did not identify a product problem.The cause of the event could not be determined.
|