MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ACTIS BROACH SZ 3; BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ACTIS BROACH SZ 3; BROACHES Back to Search Results
Catalog Number 201001030
Device Problems Fitting Problem (2183); Device-Device Incompatibility (2919); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the post on actis broaches are worn allowing trial necks to slide off post.Needing to replace three sets of actis broaches.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination of the returned instrument found the post of the broach to be scratched.The root cause is attributed to inadvertent use error.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Added: d4, concomitant medical products, device manufacture date.Device evaluated by mfr.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS BROACH SZ 3
Type of Device
BROACHES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9224688
MDR Text Key183704611
Report Number1818910-2019-111407
Device Sequence Number1
Product Code KWL
UDI-Device Identifier10603295393658
UDI-Public10603295393658
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201001030
Device Lot NumberPG276738
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2019
Initial Date Manufacturer Received 10/02/2019
Initial Date FDA Received10/23/2019
Supplement Dates Manufacturer Received11/08/2019
11/25/2019
Supplement Dates FDA Received11/08/2019
11/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age73 YR
-
-