• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JABIL CIRCUIT SINGAPORE PTE LTD BIS; ELECTRODE, CUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JABIL CIRCUIT SINGAPORE PTE LTD BIS; ELECTRODE, CUTANEOUS Back to Search Results
Model Number 186-0106
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/24/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the device could not provide correct readings.The customer indicated no patient harm.
 
Manufacturer Narrative
Evaluation summary: one device was inspected.The device failed to meet specification as it was received or made available for evaluation.The device was visually inspected and no physical anomalies were found.The device was tested with a sensor lot code tester; the device was within expiration date and displayed a ¿sensor valid¿ message.An impedance check was performed using a digital multimeter, and all electrodes passed the impedance check.The investigation found the device to function normally.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIS
Type of Device
ELECTRODE, CUTANEOUS
Manufacturer (Section D)
JABIL CIRCUIT SINGAPORE PTE LTD
1 upland rd
norwood MA 02062
MDR Report Key9236255
MDR Text Key163707293
Report Number2936999-2019-00892
Device Sequence Number1
Product Code GXY
UDI-Device Identifier20884521134307
UDI-Public20884521134307
Combination Product (y/n)N
PMA/PMN Number
K093183
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/09/2020
Device Model Number186-0106
Device Catalogue Number186-0106
Device Lot Number1209181K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2019
Initial Date Manufacturer Received 10/08/2019
Initial Date FDA Received10/24/2019
Supplement Dates Manufacturer Received12/10/2019
Supplement Dates FDA Received12/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-