• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES Back to Search Results
Model Number A22251C
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/15/2019
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device was not returned to olympus for evaluation / investigation since it was reportedly discarded by the user facility.Therefore, the exact cause of the user's experience and the reported phenomenon could not be conclusively determined and is being judged as unknown.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event / incident will be recorded for trending and surveillance purposes.
 
Event Description
Olympus was informed that during a therapeutic transurethral resection (tur) procedure, the roller ball at the distal end of the hf resection electrode broke off and fell into the patient¿s bladder.However, no fragment remained inside the patient since it was reportedly retrieved.No further information was provided but the intended procedure was completed successfully using a similar device and there was no report about an adverse event or patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key9236951
MDR Text Key165053106
Report Number9610773-2019-00146
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761036815
UDI-Public14042761036815
Combination Product (y/n)N
PMA/PMN Number
K951863
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 10/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/05/2023
Device Model NumberA22251C
Device Catalogue NumberA22251C
Device Lot Number1000030183
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/16/2019
Initial Date FDA Received10/25/2019
Supplement Dates Manufacturer Received10/25/2019
Supplement Dates FDA Received10/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-