• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 10/28/2019.
 
Event Description
The customer reported interface is broken.The device was in clinical use at the time the issue was discovered, however, there was no patient or user harm reported.
 
Manufacturer Narrative
Date received by manufacturer: 07nov2019.Date of report: 13nov2019.The service technician confirmed the interface outside the air outlet is broken and needs to be replaced.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
Date received by manufacturer: 28oct2019.Date of report: 11nov2019.The service technician confirmed the air outlet of the equipment is broken and it needs to be replaced.Pending repair.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
Mfr received date: 20 nov 2019.The service technician confirmed the joint has been replaced and the ventilator is back in normal use.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9244331
MDR Text Key170710089
Report Number2031642-2019-10271
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/15/2019
Initial Date FDA Received10/28/2019
Supplement Dates Manufacturer Received10/15/2019
10/15/2019
10/15/2019
Supplement Dates FDA Received11/11/2019
11/13/2019
11/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK AND HUMIDIFIER.
-
-