Catalog Number 10379675
Device Problem False Negative Result (1225)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/01/2019
Event Type  malfunction  
Manufacturer Narrative
The customer stated that repeat testing was performed to confirm correct results which met the clinical picture.Siemens service reviewed cleaning, maintenance and technique with the customer.Cleaning the calibration bar was not performed as often as recommended.New test tables were received and replaced and the customer stated they have not had any recent issues.
 
Event Description
The customer reported a false negative urine hcg on the clinitek status+ when compared to a quantitative serum hcg assay.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
6107052212
MDR Report Key9245057
MDR Text Key164915693
Report Number3002637618-2019-00116
Device Sequence Number1
Product Code JIL
UDI-Device Identifier00630414574639
UDI-Public00630414574639
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Physician
Type of Report Initial
Report Date 10/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/08/2019
Initial Date FDA Received10/28/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-