• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION TR BAND; CLAMP, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION TR BAND; CLAMP, VASCULAR Back to Search Results
Model Number N/A
Device Problems Inflation Problem (1310); Material Puncture/Hole (1504)
Patient Problems Hematoma (1884); Blood Loss (2597)
Event Date 10/03/2019
Event Type  Injury  
Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Healthcare professional - unknown.The actual device has been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
 
Event Description
The user facility reported that the tr band was found to have a hole in it when trying to inflate it with the syringe.Additional information was received on 15oct2019.The type of procedure that was performed being performed was a left heart catheterization.The location of the hole could not be seen by the account.The tr band was not holding any air.The patient was stable.The procedure was completed by using a different tr band.The estimated blood loss was less than 250cc.Additional information was received on 21oct2019.Per the account it appeared there was a puncture in the tr band "pillow" which subsequently led to bleeding and the patient developed a significant hematoma.
 
Manufacturer Narrative
This report is being submitted as follow up no.1 to update section h3, and to provide the completed investigation results.One used large tr band assembly was returned for product evaluation.Visual inspection revealed that the communication port which attaches the two balloons was ripped apart completely.Leak testing was attempted to be performed on the device.However, due to the communication port being fully open the balloons were not able to be inflated.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.Based on the investigation results it is likely that the balloons were attempted to be separated prior to inflating which could have caused the reported event.However, the exact cause of the reported event cannot be definitively determined based on the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TR BAND
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
MDR Report Key9261952
MDR Text Key164487922
Report Number1118880-2019-00306
Device Sequence Number1
Product Code DXC
UDI-Device Identifier00389701011356
UDI-Public00389701011356
Combination Product (y/n)N
PMA/PMN Number
K152525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model NumberN/A
Device Catalogue NumberTRB29-LRG
Device Lot NumberXC28
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2019
Initial Date Manufacturer Received 10/03/2019
Initial Date FDA Received10/31/2019
Supplement Dates Manufacturer Received11/05/2019
Supplement Dates FDA Received11/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age56 YR
Patient Weight107
-
-