RAYNHAM MICROSENSOR VENTRICULAR CATHETER KIT; ICP MICROSENSORS (ROHS) Back to Search Results
Catalog Number 826653
Device Problem No Device Output (1435)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/14/2019
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow up report will be submitted.
 
Event Description
It was reported that a microsensor ventricular catheter kit had no monitoring data showing after being connected with the icp monitor.There was no surgical delay nor adverse consequence to the patient.
 
Manufacturer Narrative
The microsensor was received for evaluation.Dhr - microsensor 82-6653, s/n (b)(6), (lot # j10t76), met specifications when released on october 9th, 2018.The release quantity was (b)(4) and the lot expires on september 30th, 2020.Failure analysis - catheter was received in a drainage tube.The device passed electronic, noise, and signal drift tests; internal calibration and linearity/hysteresis tests were omitted due to the drainage tube.The issue of the complaint has not been confirmed.Based on the failure analysis, no root cause could be determined as the device worked normally.Between 05nov2019 and 30jun2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from plainsboro, nj to princeton, nj.Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa (b)(4) and nc 20-011 have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period 05nov2019 through 30jun2020.Integra lifesciences contacted (b)(6) , director of regulatory programs, office of product evaluation and quality and (b)(6) , assistant director, mdr team, office of product evaluation and quality on july 8-9, 2020 to report these issues regarding mdr reports.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROSENSOR VENTRICULAR CATHETER KIT
Type of Device
ICP MICROSENSORS (ROHS)
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key9263280
MDR Text Key191853160
Report Number1226348-2019-00512
Device Sequence Number1
Product Code GWM
Combination Product (y/n)N
PMA/PMN Number
K991222
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number826653
Device Lot NumberJ10T76
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2020
Initial Date Manufacturer Received 10/15/2019
Initial Date FDA Received10/31/2019
Supplement Dates Manufacturer Received03/25/2020
Supplement Dates FDA Received07/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-