• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS - HOUSTON PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problems Premature End-of-Life Indicator (1480); Under-Sensing (1661); Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/14/2015
Event Type  malfunction  
Event Description
The generator was explanted and returned due to prophylactic replacement.No allegations were made against the vns generator.During product analysis, a visual assessment of the printed circuit board assembly (pcba) showed contaminates on the trimmed edge of the pcba.The contamination that was observed on the trimmed edge of the pcba suggested probable resistive electrical paths were established between the copper edges on the trimmed edge of the pcba, which contributed to the supply current and sense settings conditions.Comprehensive electrical testing showed that the generator did not perform according to functional specifications.The results of the final electrical test suggest undersensing from the heart rate sensing feature.The device history record of the generator was reviewed, and it was confirmed that the device was laser-routed during the manufacturing process.The device met all specifications for release prior to distribution.Internal investigation showed that the observed premature battery life indicator and failed heart rate undersensing were predominately caused by conductive debris from the laser-routing process resulting in leakage paths.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
rachel kohn
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key9267361
MDR Text Key164839940
Report Number1644487-2019-02109
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/09/2017
Device Model Number106
Device Lot Number4410
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2019
Event Location Other
Initial Date Manufacturer Received 10/08/2019
Initial Date FDA Received11/01/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age43 YR
-
-