Catalog Number 71170047
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/03/2019
Event Type  malfunction  
Event Description
It was reported that during surgery the device was dull.A delay of 5 minutes was reported and there was no smith and nephew backup device available, they had to use the hospital sets.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.Only one device was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The cutting edges of the device were damaged with gouges in the surface, rendering the device inoperable.The device was manufactured in 2014 and exhibits signs of extensive wear / usage.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATE CUTTING FORCEPS
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9268813
MDR Text Key164723451
Report Number1020279-2019-03856
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71170047
Device Lot Number14ATDK003
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/03/2019
Initial Date FDA Received11/01/2019
Supplement Dates Manufacturer Received04/02/2020
Supplement Dates FDA Received04/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-