• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 990063-020
Device Problems Entrapment of Device (1212); Material Integrity Problem (2978); Material Twisted/Bent (2981)
Patient Problem Tissue Damage (2104)
Event Date 10/09/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the mapping catheter was unable to hold its shape and was unable to be maneuvered.After the final confirmation for potential, the tip of the mapping catheter did not work and a resistance was felt.The sheath and balloon catheter were advanced to the mapping catheter tip to be able to remove the mapping catheter.The system was removed and tissue was found to be adhered to the tip of the mapping catheter and the loop was kinked and damaged.The case was able to be completed with cryo.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER - 20 MM
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9274235
MDR Text Key165198560
Report Number9612164-2019-04627
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/26/2021
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number217923114
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/09/2019
Initial Date FDA Received11/04/2019
Date Device Manufactured06/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-