LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Patient Involvement (2645)
Event Date 10/08/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(4).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported error message.He addressed the failure back to the shaft angle encoder.He replaced the part and the issue was solved.Functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report of a s5 double head pump shaft angle encoder error message during setup.There was no patient involvement.
 
Manufacturer Narrative
Livanova deutschland requested the replaced shaft angle encoder for investigation.As soon as the pump speed was increased, the error message reported by the customer was displayed on the control display module.The shaft angle encoder has been checked: the light diode between two pins was found to be damaged.This has been identified to be the root cause of the reported event.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key9278508
MDR Text Key195738561
Report Number9611109-2019-00868
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/08/2019
Initial Date FDA Received11/05/2019
Supplement Dates Manufacturer Received11/19/2019
Supplement Dates FDA Received12/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-